Abstract :
While performing posterior communicating aneurysm coiling, the framing of the aneurysm was performed but the coil detachment device was having an error which was only known when the coil was released into the aneurysm. The biomedical engineer within the operating room created a twelve voltage power supply circuit which was used to detach all the coils, thus rescuing the situation and saving the patient.
Keywords :
Aneurysm, Coil Detachment, Coil Release Device, Coils, Twelve Voltage Power Supply.References :
- Rajbhandari P, Neupane A, Rajbhandari S, Shrestha P, Acharya S, Shrestha R, et al. Initial experience with aneurysm coiling in Nepal. Nepal Journal of Neuroscience [Internet]. 2018 Sep 4;15(2):30–5
- Piotin M, Blanc R, Spelle L, Mounayer C, Piantino R, Schmidt PJ, et al. Stent-assisted coiling of intracranial aneurysms: Clinical and angiographic results in 216 consecutive aneurysms. Stroke [Internet]. 2010 Jan 1;41(1):110–5.
- (4) (PDF) Endovascular minimally invasive treatment of the intracranial aneurysms – first 124 cases [Internet].